Skip to content

RST Spindles

The choice of coaxial cylinder, cone and vane spindles accommodates viscosity measurement of a wide variety of materials.
Rst spindle brabender
The RST series Rheometers are available as three different models each with a different spindle geometry:

RST-CC Rheometers use coaxial cylinder spindle geometry
RST-CPS Rheometers use cone and plate spindle geometry
RST-SST Rheometers use vane spindle geometry
RST-90Y Brabender-type Spindle BU or Brabender Units spindle assembly for the RST Rheometer and is made from type 304 stainless steel. It can be used for ASTM Standard Method C474, for testing joint compounds

rst group
  • Specifications +


    SPECIFICATIONS

    RST Spindle RS Rheometers with serial numbers beginning with “302” and “303” use different spindles. Call for details.
    SPINDLE VISCOSITY RANGE SHEAR RATE MAX. SHEAR STRESS SAMPLE VOLUME
    COAXIAL Pa•s sec-1 Pa mL
    CCT-DG 0.00005-4.07K 0.043-5.64K 177 15.7
    CCT-40 0.0003-27.6K 0.0215-2.79K 594 68.5
    CCT-25 0.002-177K 0.013-1.67K 2.28K 16.8
    CCT-14 0.012-1M 0.013-1.68K 13K 3.4
    CCT-8 0.065-5.41M 0.013-1.672K 69.6K 1.0
    CONE Pa•s sec-1 Pa mL
    RCT-25-1 0.005-407K 0.06-7.8K 24.4K 0.1
    RCT-25-2 0.01-814K 0.03-3.9K 24.4K 0.2
    RCT-50-1 0.0006-50.9K 0.06-7.8K 3.05K 1.0
    RCT-50-2 0.0012-101K 0.03-3.9K 3.05K 2.0
    RCT-75-1* 0.0002-15K 0.06-7.8K 905 2.5
    RCT-75-2* 0.0004-30K 0.03-3.9K 905 5.0
    PLATE Pa•s sec-1 Pa mL
    RPT-25 0.03-2.49M 0.013-1.7K 32.6K 0.5
    RPT-50 0.002-155K 0.027-3.4K 4.07K 2.0
    RPT-75* 0.0004-30.7K 0.04-5.1K 1.2K 4.5

    VANE
    SPINDLE
    VANE LENGTH (mm) VANE DIAMETER
    (mm)
    SHEAR STRESS
    (Pa)
    VT-10-5 10 5 330-210K
    VT-20-10 20 10 41-27K
    VT-20-20 20 20 9-5.9K
    VT-30-15 30 15 12-8K
    VT-40-20 40 20 5.2-3.4K
    VT-40-40 40 40 1.2-740
    VT-50-25 50 25 2.7-1.7K
    VT-60-8 60 8 24-15K
    VT-60-15 60 15 7-4.3K
    VT-60-30 60 30 1.6-1K
    VT-80-40 80 40 0.7-420
    VT-80-70 80 70 0.2-120
    *For use with water bath version only. 1 Pa · s = 1,000 cP K = 1 thousand M = 1 million
    Note: 1. Values based on minimum speed of 1 RPM and maximum speed of 1000 RPM
              2. 75 mm plates cannot be used with Peltier Plate or electrically heated rheometers
For Immediate assistance,
Call 1-508-946-6200 (US ET)
Monday - Friday from 8:00AM - 5:00PM