Skip to content

VPXL and VPXL Autosampler Manual Glossary

  • AC Power Connection
    Power supply requirement for the Vapor Pro XL

  • Autosampler
    Optional accessory allowing automated testing of up to 16 samples sequentially

  • Balance
    External Accessory used to weigh samples before testing

  • Bar Code Reader
    Optional accessory for entering data like sample IDs and lot numbers by scanning barcodes, useful for efficiency and accuracy

  • Calibration
    Process of verifying or adjusting the instrument’s accuracy. Includes RH system calibration, temperature calibration, and flow sensor calibration

  • Carrier Gas
    Either dry nitrogen or dry air is used in the instrument to transport volatiles to the sensor. Recommended to have a dew point of -40°C or lower
     
  • Cold Start
    Featuring allowing the oven to start a test at room temperature before ramping up to the test setpoint, usually to enhance analysis for certain samples

  • Dry Air Generator
    Accessory that provides a continuous supply of dry air to the instrument without needing a compressed air source

  • Dry Gas Inlet
    Connection point on the instrument where the dry gas (air or nitrogen) is supplied for testing

  • Ethernet Setup
    Configures the device’s Ethernet connection, allowing remote access via a local network

  • End Criteria
    Determines when a test end based on rate, time, or prediction settings to meet specific analysis requirements

  • Factory Idle
    Pre-programmed setting used to rapidly cool the instrument’s oven

  • Flow Check and Adjustments
    Procedure to verify and adjust the carrier gas flow rate to ensure consistent testing conditions

  • Graph Data
    Graphical representation of test results showing moisture loss rate and temperature against time

  • Hydrophobic Filter Kit
    Accessory recommended for samples with unreacted or volatile components to protect the RH sensor from contamination

  • Insert for Crimp Top Sample Vials
    Inserts are available to accommodate crimp-top vials in various sizes for specific sample requirements

  • Karl Fischer Correlation
    Instrument provides a moisture analysis method that correlates with the Karl Fischer titration without using reagents

  • Login Setup
    Security features restricting access to authorized users by requiring login credentials for critical functions like calibration and configuration

  • Memory
    Storage capacity of the instrument, holding up to 20 test programs and 1000 test results

  • Nitrogen Control c
    Feature for controlling the nitrogen flow, useful for conserving nitrogen when the instrument is idle

  • Oven Fan
    Fan in the oven that speeds up cooling between tests to reduce waiting times and improve efficiency

  • Pass/Fail Criteria
    Option to set moisture limits for tests, where the results can be categorized as “pass” or “fail” based on the predefined moisture thresholds

  • Pre-Purge
    Setting available in the Autosampler to remove headspace moisture from vials before testing, useful for sensitive samples

  • Prediction Algorithm
    Feature that estimates the final moisture results before all moisture is driven off, thus reducing test times

  • RH Sensor
    (Arching) Relative humidity sensor used in the Vapor Pro XL to measure the moisture content of the sample

  • Sample Prompt
    Custom message displayed at the start of a test to instruct the user on sample preparation

  • Sample Weight Entry
    Allows users to input sample weights manually or by connecting a balance for automatic input

  • Standby Timeout
    Features that reduce the oven temperature to a standby level after a specified idle time to conserve energy and prolong component life

  • System Calibration
    Calibration ensures the accuracy of the moisture measurement system. Should be done regularly and before the first test

  • Test Program
    Set of predefined parameters for a specific sample type, including sample size, heater temperature, and test criteria. Up to 250 programs can be saved

  • Temperature Calibration
    Ensures the accuracy of the oven temperature, done with a Temperature Calibration Interface (TCI) kit

  • Time Rate
    Test ending criterion where the test ends after a specified rate is achieved following a set duration

  • USB Host
    USB port for connecting accessories like a printer, keyboard, or USB drive

  • USB Slave
    Port for connecting the instrument to a PC for data transfer or control

  • Web Server
    Optional Feature allowing remote access to the instrument’s data and configuration through a local network or intranet

  • Zero Vial Purge
    Setting to capture all sample moisture by setting the vial purge time to zero, often used when residual moisture needs to be retained

  • Shear Stress
    Stress due to the Torsional Load. It is the stress between parallel layers of powder (Flow Function and Time Consolidated Flow Function tests) or a layer of powder and the lid (Wall Friction test)

  • Steady State Flow
    Continuous powder flow where the Normal and Shear Stresses have reached a constant level

  • Time Consolidation Test
    Compacts the powder at a given Consolidation Level for an extended period (hours, overnight to represent static storage in a hopper) before measuring the Failure Locus

  • Torsional Distance
    Distance that the trough has rotated from its starting position at the beginning of the test

  • Torsional Load
    Rotational force applied during testing to simulate shear conditions within the powder sample

  • Trough
    Container that holds the powder sample during testing, available in standard and small volume sizes

  • Unconfined Failure Strength
    Maximum stress a compacted powder can withstand at a free surface, after it has been compacted to a given Consolidation Level

  • Wall Cohesion
    Measure of the adhesives force between a powder and the wall of a hopper or chute

  • Wall Friction Test
    Test to determine the interaction between powder and wall surfaces, influencing hoper design